Form 8-K

 

 

UNITED STATES

SECURITIES AND EXCHANGE COMMISSION

Washington, D.C. 20549

 

 

FORM 8-K

 

 

CURRENT REPORT

Pursuant to Section 13 or 15(d) of the Securities Exchange Act of 1934

Date of Report (Date of earliest event reported): January 20, 2014

 

 

Tandem Diabetes Care, Inc.

(Exact name of registrant as specified in its charter)

 

 

Delaware

(State or other jurisdiction of incorporation)

 

001-36189   20-4327508
(Commission File Number)   (IRS Employer Identification No.)
11045 Roselle Street, San Diego, CA   92121
(Address of principal executive offices)   (Zip code)

Registrant’s telephone number, including area code (858) 366-6900

N/A

(Former name or former address, if changed since last report)

 

 

Check the appropriate box below if the Form 8-K filing is intended to simultaneously satisfy the filing obligation of the registrant under any of the following provisions (see General Instruction A.2. below):

 

¨ Written communications pursuant to Rule 425 under the Securities Act (17 CFR 230.425)

 

¨ Soliciting material pursuant to Rule 14a-12 under the Exchange Act (17 CFR 240.14a-12)

 

¨ Pre-commencement communications pursuant to Rule 14d-2(b) under the Exchange Act (17 CFR 240.14d-2(b))

 

¨ Pre-commencement communications pursuant to Rule 13e-4(c) under the Exchange Act (17 CFR 240.13e-4(c))

 

 

 


Item 8.01 Other Events.

On January 20, 2014 we announced the expansion of a previously announced voluntary recall of specific lots of insulin cartridges that are used with the t:slim® Insulin Pump. A press release announcing the expansion of the voluntary recall is attached hereto as Exhibit 99.1 and is incorporated herein by reference.

Item 9.01 Financial Statements and Exhibits.

 

(d) Exhibits.

 

Exhibit No.

  

Description

99.1    Press Release dated January 20, 2014


SIGNATURES

Pursuant to the requirements of the Securities Exchange Act of 1934, the registrant has duly caused this report to be signed on its behalf by the undersigned hereunto duly authorized.

 

Date: January 21, 2014

      Tandem Diabetes Care, Inc.
     

 /s/ David B. Berger

     

David B. Berger, General Counsel


EXHIBIT INDEX

 

Exhibit No.

  

Description

99.1    Press Release dated January 20, 2014

 

EX-99.1

Exhibit 99.1

 

LOGO

   Media Contact:

Steve Sabicer

714-907-6264

ssabicer@thesabicergroup.com

 

Investor Contact:

Susan Morrison

858-366-6900 x7005

smorrison@tandemdiabetes.com

  

FOR IMMEDIATE RELEASE

Tandem Diabetes Care Expands Voluntary Recall for Select Lots of Insulin Cartridges Used with t:slim Insulin Pump

Company Is Providing Replacement Cartridges to Its Customers at No Additional Charge

SAN DIEGO, January 20, 2014 — Tandem Diabetes Care®, Inc. (NASDAQ: TNDM), or Tandem, today announced that it is expanding a voluntary recall of specific lots of insulin cartridges that are used with the t:slim® Insulin Pump. The affected cartridges may be at risk for leaking. This could potentially result in the device delivering too much or too little insulin, which can lead to a serious adverse event.

“The safety of our customers is our top priority. For that reason, even after our previous announcement Tandem’s team continued to perform further testing. Based on this work we recognized that additional cartridge lots could also be at risk for leaking and so we are expanding the scope of this recall,” said Kim Blickenstaff, President and CEO of Tandem. “We are confident that we have identified all of the affected cartridge lots that may have been shipped to customers or distributors and that we have implemented appropriate corrective actions to prevent this from happening in the future.”

Customers should discontinue using cartridges labeled with any of the lot numbers listed below. The affected lots shipped to customers or distributors represent an aggregate of approximately 13,000 boxes of cartridges (10 cartridges per box). The affected lots announced on January 10, 2014 were shipped from December 17, 2013 to January 10, 2014. The affected lots announced on January 20, 2014 were shipped prior to October 16, 2013.

Affected Lot Numbers

 

Announced January 20, 2014

 

Announced January 10, 2014

M000857

  M001414   M001454   M001963   M002028

M000869

  M001415   M001455   M001964   M002029

M001344

  M001416   M001456   M001973   M002030

M001345

  M001417   M001457   M001974   M002082

M001346

  M001420   M001458   M001979   M002083

M001347

  M001421   M001459   M001980   M002096

M001389

  M001422   M001460   M001987   M002097

M001390

  M001423   M001528   M001988   M002099

M001391

  M001451   M001529   M001990   M002100

M001392

  M001452   M001530   M001991   M002119

M001393

  M001453   M001532   M002027   M002120


All other cartridge lots shipped to customers or distributors and the t:slim Insulin Pump are not affected by this recall. The Company, or its authorized distributors, is asking customers who received affected cartridges to call Tandem Technical Support to receive replacement cartridges at no charge. Tandem expects to have sufficient quantities of cartridges to replace affected lots in a timely manner. Tandem Customer Support is available 24 hours a day, 7 days a week at 1-877-801-6901.

The Company has determined that under certain conditions a specific piece of equipment used in the cartridge test process caused the affected cartridges to be at risk for leaking. Tandem has already implemented modifications to the cartridge manufacturing test process to prevent this issue from occurring in the future, and the Company’s manufacturing facility is able to support ongoing customer demand.

The Company is working with its customers and distributors to replace cartridges subject to the recall. Based on the periods during which the affected lots were shipped, the Company believes a significant number of the affected cartridges have already been used and therefore will not be exchanged. In addition to the expenses associated with replacing affected cartridges, the Company may also incur additional logistical shipping fees and other recall coordination costs. Tandem is also in the process of evaluating additional cartridges still in the Company’s internal inventory to determine whether they will be usable, and may incur expenses with those that are unusable.

About Tandem Diabetes Care, Inc.

Tandem Diabetes Care, Inc. is a medical device company with an innovative, user-centric and integrated approach to the design, development and commercialization of products for people with insulin-dependent diabetes. The Company manufactures and sells the t:slim Insulin Pump, the slimmest and smallest durable insulin pump currently on the market, and the first and only with a high resolution, color touch screen. Tandem is based in San Diego, Calif.

Forward Looking Statement

This press release includes forward-looking statements concerning Tandem’s voluntary recall of specific lots of cartridges that are used with the t:slim® Insulin Pump, the scope of the recall and the Company’s belief that it has identified all of the affected cartridge lots that may have been shipped to customers or distributors, the cause of the potential cartridge leak, the implementation of remedial actions that are intended to prevent this issue from occurring in the future, Tandem’s ability to replace affected cartridges on a timely basis, the Company’s ability to continue to manufacture cartridges without significant disruption in order to meet ongoing customer demand, Tandem’s estimate of the portion of affected cartridges that will be replaced and estimated expenses and charges that Tandem may incur as a result of the recall, or from the write off of internal inventory. Tandem’s actual results may differ materially from those indicated in these forward looking statements due to risks and uncertainties, including: future actions of the FDA or any other regulatory body or governmental authority; Tandem’s ability to modify its cartridge manufacturing process in order to prevent this issue from occurring in the future; Tandem’s reliance on distributors to facilitate the product recall; the potential that Tandem may subsequently discover additional lots of cartridges subject to this recall; the possibility that the actual replacement rate for affected cartridges may be higher or lower than estimated; the potential that the company may incur additional expenses that it does not anticipate at this time; the inherent difficulty of predicting the total financial impact associated with the recall; the possibility that this recall could subject Tandem to claims or proceedings which may adversely impact its business and financial condition; and other risks identified in Tandem’s Registration Statement on Form S-1. Investors are cautioned not to place undue reliance on these forward-looking statements, which speak only as of the date of this release. Tandem undertakes no obligation to update any forward-looking statement in this press release.

# # #